SU‐GG‐T‐362: Is Distance‐Dependent Energy Correction Needed for Dosimetric Measurements Surrounding Brachytherapy Sources?
Purpose: The answer depends on both the dosimeter's energy response and energy spectrum at distance (point of measurement) from the source. The latter is object of this study. Method and Materials: The spectra in water were calculated with Monte Carlo code MCNP5. The source point was centered within a spherical water medium, divided into concentric spherical shells of various thickness. The f4 fluence tally of MCNP, with fine energy bins, was used to register photon energy spectrum in thin shells (0.01 cm thick) centered at radii 0.2, 0.5, 1, 2, 3, …9 and 10 cm. One million particles were used for each MCNP run, to achieve good statistical uncertainty. Seven commonly used brachytherapy sources were studied: 60Co (E