A systematic assessment and optimization of photon-counting CT for lung density quantifications.

Publication ,  Journal Article
Sotoudeh-Paima, S; Segars, WP; Ghosh, D; Luo, S; Samei, E; Abadi, E
Published in: Med Phys
April 2024

BACKGROUND: Photon-counting computed tomography (PCCT) has recently emerged into clinical use; however, its optimum imaging protocols and added benefits remains unknown in terms of providing more accurate lung density quantification compared to energy-integrating computed tomography (EICT) scanners. PURPOSE: To systematically assess the performance of a clinical PCCT scanner for lung density quantifications and compare it against EICT. METHODS: This cross-sectional study involved a retrospective analysis of subjects scanned (August-December 2021) using a clinical PCCT system. The influence of altering reconstruction parameters was studied (reconstruction kernel, pixel size, slice thickness). A virtual CT dataset of anthropomorphic virtual subjects was acquired to demonstrate the correspondence of findings to clinical dataset, and to perform systematic imaging experiments, not possible using human subjects. The virtual subjects were imaged using a validated, scanner-specific CT simulator of a PCCT and two EICT (defined as EICT A and B) scanners. The images were evaluated using mean absolute error (MAE) of lung and emphysema density against their corresponding ground truth. RESULTS: Clinical and virtual PCCT datasets showed similar trends, with sharper kernels and smaller voxel sizes increasing percentage of low-attenuation areas below -950 HU (LAA-950) by up to 15.7 ± 6.9% and 11.8 ± 5.5%, respectively. Under the conditions studied, higher doses, thinner slices, smaller pixel sizes, iterative reconstructions, and quantitative kernels with medium sharpness resulted in lower lung MAE values. While using these settings for PCCT, changes in the dose level (13 to 1.3 mGy), slice thickness (0.4 to 1.5 mm), pixel size (0.49 to 0.98 mm), reconstruction technique (70 keV-VMI to wFBP), and kernel (Qr48 to Qr60) increased lung MAE by 15.3 ± 2.0, 1.4 ± 0.6, 2.2 ± 0.3, 4.2 ± 0.8, and 9.1 ± 1.6 HU, respectively. At the optimum settings identified per scanner, PCCT images exhibited lower lung and emphysema MAE than those of EICT scanners (by 2.6 ± 1.0 and 9.6 ± 3.4 HU, compared to EICT A, and by 4.8 ± 0.8 and 7.4 ± 2.3 HU, compared to EICT B). The accuracy of lung density measurements was correlated with subjects' mean lung density (p < 0.05), measured by PCCT at optimum setting under the conditions studied. CONCLUSION: Photon-counting CT demonstrated superior performance in density quantifications, with its influences of imaging parameters in line with energy-integrating CT scanners. The technology offers improvement in lung quantifications, thus demonstrating potential toward more objective assessment of respiratory conditions.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Med Phys

DOI

EISSN

2473-4209

Publication Date

April 2024

Volume

51

Issue

4

Start / End Page

2893 / 2904

Location

United States

Related Subject Headings

  • Tomography, X-Ray Computed
  • Retrospective Studies
  • Pulmonary Emphysema
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Lung Diseases
  • Lung
  • Humans
  • Emphysema
  • Cross-Sectional Studies
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Sotoudeh-Paima, S., Segars, W. P., Ghosh, D., Luo, S., Samei, E., & Abadi, E. (2024). A systematic assessment and optimization of photon-counting CT for lung density quantifications. Med Phys, 51(4), 2893–2904. https://doi.org/10.1002/mp.16987
Sotoudeh-Paima, Saman, W Paul Segars, Dhrubajyoti Ghosh, Sheng Luo, Ehsan Samei, and Ehsan Abadi. “A systematic assessment and optimization of photon-counting CT for lung density quantifications.Med Phys 51, no. 4 (April 2024): 2893–2904. https://doi.org/10.1002/mp.16987.
Sotoudeh-Paima S, Segars WP, Ghosh D, Luo S, Samei E, Abadi E. A systematic assessment and optimization of photon-counting CT for lung density quantifications. Med Phys. 2024 Apr;51(4):2893–904.
Sotoudeh-Paima, Saman, et al. “A systematic assessment and optimization of photon-counting CT for lung density quantifications.Med Phys, vol. 51, no. 4, Apr. 2024, pp. 2893–904. Pubmed, doi:10.1002/mp.16987.
Sotoudeh-Paima S, Segars WP, Ghosh D, Luo S, Samei E, Abadi E. A systematic assessment and optimization of photon-counting CT for lung density quantifications. Med Phys. 2024 Apr;51(4):2893–2904.

Published In

Med Phys

DOI

EISSN

2473-4209

Publication Date

April 2024

Volume

51

Issue

4

Start / End Page

2893 / 2904

Location

United States

Related Subject Headings

  • Tomography, X-Ray Computed
  • Retrospective Studies
  • Pulmonary Emphysema
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Lung Diseases
  • Lung
  • Humans
  • Emphysema
  • Cross-Sectional Studies