Quantitative scatter measurement in digital radiography using a photostimulable phosphor imaging system.

Publication ,  Journal Article
Floyd, CE; Lo, JY; Chotas, HG; Ravin, CE
Published in: Med Phys
1991

X-ray scatter fractions measured with two detectors are compared: a photostimulable phosphor system (PSP) and a conventional film-screen technique. For both detection methods, a beam-stop technique was used to estimate the scatter fraction in polystyrene phantoms. These scatter fraction measurements are compared to previously reported film-based measurements. Scatter fractions obtained with the PSP were in good agreement both with measurements using film as well as with most previously reported measurements. For the PSP measurements, repeatability was better than 1%. It was found that the PSP provides a precise x-ray detector for quantitative scatter measurement in digital radiography.

Duke Scholars

Published In

Med Phys

DOI

ISSN

0094-2405

Publication Date

1991

Volume

18

Issue

3

Start / End Page

408 / 413

Location

United States

Related Subject Headings

  • X-Ray Intensifying Screens
  • Scattering, Radiation
  • Polystyrenes
  • Nuclear Medicine & Medical Imaging
  • Models, Structural
  • Luminescent Measurements
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering
  • 1112 Oncology and Carcinogenesis
  • 0903 Biomedical Engineering
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Floyd, C. E., Lo, J. Y., Chotas, H. G., & Ravin, C. E. (1991). Quantitative scatter measurement in digital radiography using a photostimulable phosphor imaging system. Med Phys, 18(3), 408–413. https://doi.org/10.1118/1.596687
Floyd, C. E., J. Y. Lo, H. G. Chotas, and C. E. Ravin. “Quantitative scatter measurement in digital radiography using a photostimulable phosphor imaging system.Med Phys 18, no. 3 (1991): 408–13. https://doi.org/10.1118/1.596687.
Floyd, C. E., et al. “Quantitative scatter measurement in digital radiography using a photostimulable phosphor imaging system.Med Phys, vol. 18, no. 3, 1991, pp. 408–13. Pubmed, doi:10.1118/1.596687.
Floyd CE, Lo JY, Chotas HG, Ravin CE. Quantitative scatter measurement in digital radiography using a photostimulable phosphor imaging system. Med Phys. 1991;18(3):408–413.

Published In

Med Phys

DOI

ISSN

0094-2405

Publication Date

1991

Volume

18

Issue

3

Start / End Page

408 / 413

Location

United States

Related Subject Headings

  • X-Ray Intensifying Screens
  • Scattering, Radiation
  • Polystyrenes
  • Nuclear Medicine & Medical Imaging
  • Models, Structural
  • Luminescent Measurements
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering
  • 1112 Oncology and Carcinogenesis
  • 0903 Biomedical Engineering