PURPOSE: High-precision intensity-modulated radiotherapy demands high patient positioning accuracy. On-board digital tomosynthesis (DTS) provides three-dimensional (3D) image guidance for daily positioning with a lower imaging dose, faster acquisition, and more geometric flexibility than 3D cone-beam computed tomography (CBCT). This clinical study evaluated DTS as a daily imaging technique for patient positioning and compared the results with 3D CBCT and two-dimensional (2D) radiography. METHODS AND MATERIALS: Head and neck cancer patients undergoing intensity-modulated radiotherapy were studied. For each session, the patient was positioned using laser marks. On-board imaging data sets, including 2D kilovoltage radiographs, DTS, and CBCT, were obtained to measure the daily patient positioning variations. The mean and standard deviations of the positioning variations in the translational and rotational directions were calculated. The positioning differences among 2D radiography, DTS, and CBCT were analyzed. RESULTS: Image data sets were collected from 65 treatment fractions for 10 patients. The systematic patient positioning variation was <0.10 cm and 1.0 degrees one dimensionally. The random variations were 0.27-0.34 cm in the translational and 0.93 degrees -1.99 degrees in the rotational direction. The mean vector isocenter variation was 0.48 cm. DTS with 40 degrees and 20 degrees scan angles in the coronal or sagittal directions yielded the same results for patient positioning. DTS performance was comparable to that of CBCT, with positioning differences of <0.1 cm and 0.5 degrees . The positioning difference between 2D radiography and DTS was approximately 0.1 cm and 0.2 cm in the vertical/longitudinal and lateral directions. CONCLUSION: Our results have demonstrated that DTS is a comparable 3D imaging technique to CBCT for daily patient positioning of head-and-neck patients as determined by manual registration of bony anatomy.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Int J Radiat Oncol Biol Phys

DOI

ISSN

0360-3016

Publication Date

October 1, 2007

Volume

69

Issue

2

Start / End Page

598 / 606

Location

United States

Related Subject Headings

  • Tomography, X-Ray Computed
  • Technology, Radiologic
  • Radiotherapy, Intensity-Modulated
  • Oncology & Carcinogenesis
  • Imaging, Three-Dimensional
  • Image Processing, Computer-Assisted
  • Humans
  • Head and Neck Neoplasms
  • Equipment Design
  • Algorithms
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Wu, Q. J., Godfrey, D. J., Wang, Z., Zhang, J., Zhou, S., Yoo, S., … Yin, F.-F. (2007). On-board patient positioning for head-and-neck IMRT: comparing digital tomosynthesis to kilovoltage radiography and cone-beam computed tomography. Int J Radiat Oncol Biol Phys, 69(2), 598–606. https://doi.org/10.1016/j.ijrobp.2007.05.045
Wu, Q Jackie, Devon J. Godfrey, Zhiheng Wang, Junan Zhang, Sumin Zhou, Sua Yoo, David M. Brizel, and Fang-Fang Yin. “On-board patient positioning for head-and-neck IMRT: comparing digital tomosynthesis to kilovoltage radiography and cone-beam computed tomography.Int J Radiat Oncol Biol Phys 69, no. 2 (October 1, 2007): 598–606. https://doi.org/10.1016/j.ijrobp.2007.05.045.
Wu QJ, Godfrey DJ, Wang Z, Zhang J, Zhou S, Yoo S, et al. On-board patient positioning for head-and-neck IMRT: comparing digital tomosynthesis to kilovoltage radiography and cone-beam computed tomography. Int J Radiat Oncol Biol Phys. 2007 Oct 1;69(2):598–606.
Wu, Q. Jackie, et al. “On-board patient positioning for head-and-neck IMRT: comparing digital tomosynthesis to kilovoltage radiography and cone-beam computed tomography.Int J Radiat Oncol Biol Phys, vol. 69, no. 2, Oct. 2007, pp. 598–606. Pubmed, doi:10.1016/j.ijrobp.2007.05.045.
Wu QJ, Godfrey DJ, Wang Z, Zhang J, Zhou S, Yoo S, Brizel DM, Yin F-F. On-board patient positioning for head-and-neck IMRT: comparing digital tomosynthesis to kilovoltage radiography and cone-beam computed tomography. Int J Radiat Oncol Biol Phys. 2007 Oct 1;69(2):598–606.
Journal cover image

Published In

Int J Radiat Oncol Biol Phys

DOI

ISSN

0360-3016

Publication Date

October 1, 2007

Volume

69

Issue

2

Start / End Page

598 / 606

Location

United States

Related Subject Headings

  • Tomography, X-Ray Computed
  • Technology, Radiologic
  • Radiotherapy, Intensity-Modulated
  • Oncology & Carcinogenesis
  • Imaging, Three-Dimensional
  • Image Processing, Computer-Assisted
  • Humans
  • Head and Neck Neoplasms
  • Equipment Design
  • Algorithms