Errors in calculating the epicardial potential gradient field from unipolar voltage measurements during cardiac defibrillation
Publication
, Journal Article
Idriss, SF; Ideker, RE; Wolf, PD; Smith, WM
Published in: Proceedings of the Annual Conference on Engineering in Medicine and Biology
December 1, 1991
An examination is made of the errors generated when calculating the magnitude of the epicardial potential gradient using voltages recorded from three unipolar electrodes in a fixed triangular geometry on the epicardial surface. The error sensitivity of the gradient calculation increases significantly with decreasing electrode separation and increasing common-mode voltage. For example, a mean error in calculated gradient magnitude of 67% results when measuring a 5 V/cm gradient at 50 V common-mode with 2 mm interelectrode spacing and random 1% voltage error. Optimization of interelectrode spacing and common-mode voltage to reduce overall error is essential for accurate measurement of low magnitude epicardial gradients.
Duke Scholars
Published In
Proceedings of the Annual Conference on Engineering in Medicine and Biology
ISSN
0589-1019
Publication Date
December 1, 1991
Volume
13
Issue
pt 2
Start / End Page
692 / 693
Citation
APA
Chicago
ICMJE
MLA
NLM
Idriss, S. F., Ideker, R. E., Wolf, P. D., & Smith, W. M. (1991). Errors in calculating the epicardial potential gradient field from unipolar voltage measurements during cardiac defibrillation. Proceedings of the Annual Conference on Engineering in Medicine and Biology, 13(pt 2), 692–693.
Idriss, S. F., R. E. Ideker, P. D. Wolf, and W. M. Smith. “Errors in calculating the epicardial potential gradient field from unipolar voltage measurements during cardiac defibrillation.” Proceedings of the Annual Conference on Engineering in Medicine and Biology 13, no. pt 2 (December 1, 1991): 692–93.
Idriss SF, Ideker RE, Wolf PD, Smith WM. Errors in calculating the epicardial potential gradient field from unipolar voltage measurements during cardiac defibrillation. Proceedings of the Annual Conference on Engineering in Medicine and Biology. 1991 Dec 1;13(pt 2):692–3.
Idriss, S. F., et al. “Errors in calculating the epicardial potential gradient field from unipolar voltage measurements during cardiac defibrillation.” Proceedings of the Annual Conference on Engineering in Medicine and Biology, vol. 13, no. pt 2, Dec. 1991, pp. 692–93.
Idriss SF, Ideker RE, Wolf PD, Smith WM. Errors in calculating the epicardial potential gradient field from unipolar voltage measurements during cardiac defibrillation. Proceedings of the Annual Conference on Engineering in Medicine and Biology. 1991 Dec 1;13(pt 2):692–693.
Published In
Proceedings of the Annual Conference on Engineering in Medicine and Biology
ISSN
0589-1019
Publication Date
December 1, 1991
Volume
13
Issue
pt 2
Start / End Page
692 / 693